Bed Head Panels

High Pressure Manifolds / Control Panels

Oxygen/Nitrous Oxide/CO2/ another gases Manifold made with High pressure Copper pipe header with Brass Blocks to fit the High pressure Valve /NRVS as per customer choice, connected with high pressure copper pipe pig tail pipes to connect one side to High pressure Valve /NRVS and other side to cylinders. The manifold has been tested to pressure of 210kg/cm2 before despatching/installation.

Control Panels

SEMI AUTO CONTROL CONTRO PANEL/FULLY AUTO MATIC CONTROL PANELS
Product Details
Fully Automatic Change Over Unit''(Fully Automatic Control Panel) to provide reliable, uninterrupted supply of gas to a hospital or other medical facility.
The System Consist of :
The control panel connecting two Manifold Head (Left Bank and Right of cylinders) Option of gas specific Copper pigtail or SS braiding high pressure flexible Tail Pipe for connecting compressed gas cylinder to manifold though HP non-return valve to form a gas bank. Change over unit with choice of Automatic or Semi Automatic Bank switching Relief Valve to bleed pressure built up excess of 5.2 bar (75 psi) for safety.
FEATURES
  • Relief Valve to bleed pressure built up excess of 5.2 bar (75 psi) for safety
  • Dual line pressure regulator
  • Flow control regulator
  • Change over Valve
  • Power needed 230 V AC
  • Wall mounting bracket attached with the panel for easy installation
SPECIAL FEATURES
  • The automatic system fails, the machine will work in manual mode
  • In case the power supply is off, the machine will work as usual; only the digital display will not function, but the analogue display will work normally